Abstract:
A photon-photon coincidence system was
constructed for the standardization of (125)I and (109)Cd in
PTKMR-BATAN, Indonesia. Two NaI(Tl) detectors of 76 mm diameter × 6mm
thickness with 0.5mm aluminum window were used, which were positioned
approximately symmetrically to the source holder.
The electronic chain
was almost the same as for a 4πβ-γ system. The CANBERRA Multiport II
multi channel analyzer was used for energy calibration and a Philips
type PM3092 oscilloscope for visualization of the pulses. A polyethylene
plastic was used as the source substrate for the (125)I and (109)Cd
samples. The activity of a (125)I solution was measured by the
photon-photon coincidence and the efficiency extrapolation method
(Schrader and Walz, 1987), whereas the activity of a (109)Cd solution
was determined by a tracer method using (125)I (Schrader, 2006). The
result of the (125)I activity showed good agreement with the result of
measurements using a calibrated ionization chamber, and the result of
(109)Cd also showed good agreement with the measurements result using a
LEGe detector.
Pujadi Marsoem, Gatot Wurdiyanto, Hermawan Candra